FAQ
01
Without on-site installation by engineers
Please read the user manual

If the machine is received through express delivery or logistics,  there is a particularly loud exhaust sound after startup, meanwhile a concentration alarm appears a few minutes later, it is likely that the machine has been damaged during transportation. Please contact the dealer.

More+
Click to zoom out
02
The screen display of the startup device is normal
And the compressor stops, starts, or does not start

A: Please check the voltage at home. In rural areas or some old urban communities, this low voltage condition often occurs. The solution is to add a household voltage regulator for voltage stabilization. Please don't think that if the TV refrigerator works properly, your home voltage is fine!

B: If there is no problem mentioned above,please contact the after-sales personnel.

More+
Click to zoom out
03
Lacking of oxygen or low oxygen flow during
oxygen concentrator working process

A: The most majority case is caused by air leakage from the humidification bottle that has not been tightened. Please tighten the humidification bottle.

B: Check if there is scale in the humidification bottle. If there is, soak it in white vinegar and warm water.

C: If there is no above-mentioned problem, please contact the after-sales personnel.

More+
Click to zoom out
04
Temperature alarm
or E-1 prompt appears

A: Firstly, check the ventilation and heat dissipation of the machine: check if there are carpets or soft fabrics underneath the machine; Check the distance between the machine and the wall or furniture: Check if the intake filter cotton has not been cleaned for a long time.

B: If there is no above-mentioned problem, the fan may malfunction. Please contact the relevant after-sales personnel.

More+
Click to zoom out
05
Concentration alarm or E-3 prompt appears

A: If the machine with the nebulization function:The alarm is normal during the nebulization process, and it will disappear when the oxygen inhalation is completed. then oxygen inhalation starts.

B: The machine with a rotary flow meter:Check the flow rate, if it exceeds the rated flow rate, an alarm will appear. For example, if the flow rate of the AE-5 oxygen concentrator exceeds 5 liters, a concentration alarm will appear. Turn down the flow rate and the alarm will disappear.

C: Please check the voltage, especially in old residential areas, in rural areas or some cities. If the voltage of the power supply network is too low, a concentration alarm will appear.

D: If there is no above-mentioned problem, please contact after-sales service personnel.


More+
Click to zoom out
06
Power failure alarm: The machine does not run after being turned on,
the screen does not light up and there is a beeping sound.

A: Check if there is power supply at home. (Suggest changing several sockets to try)

B: Check the fuse and replace it with a new one if it blows. If it is the AE series, press the manual reset circuit breaker next to the power switch.

C: If there is no above-mentioned problem, please contact the after-sales personnel.